:TRIGger[:SEQuence][:IMMediate]
Description
Generates a trigger regardless of the trigger source selection. The command is finished before the measurement is completed and can NOT be awaited with an OPC command.
If the trigger is NOT in the Waiting for Trigger
state, the command has no effect and a Trigger Ignored
error (nr. -211
) is generated. The only difference between this command and the :TRIGger[:SEQuence]:SINGle is that this command CAN NOT be awaited and the :TRIGger[:SEQuence]:SINGle command CAN be awaited.
For more information about the trigger system, please refer to Trigger System.
Syntax | Parameter |
---|---|
:TRIGger[:SEQuence][:IMMediate] |
none |
Parameters | accepted format | allowed suffixes | possible Errors | description |
---|---|---|---|---|
none |
none |
none |
Trigger Ignored - -211 |
none |
Sample usage
Start a single trigger cycle.