:SENSe:CORRection:USERrange:DATA:FREQuencies
Description
Gets the actual user-range calibration frequencies. If no data is available (no calibration was performed) a Query
error (nr. -400
) is generated.
For more information about the calibration system please refer to Calibration.
Syntax | Response |
---|---|
:SENSe<ch>:CORRection:USERrange:DATA:FREQuencies? |
<numeric array > |
Parameters | Accepted format | Description |
---|---|---|
<ch > |
{ 1 | empty } |
Currently only one channel is supported. Empty means 1 is used as default. |
Response value | Description |
---|---|
<numeric array > |
List of frequencies in Hz used for performing the user-range calibration separated by a comma (, ). The frequencies are directly related to the frequency settings Start Frequency, Stop Frequency, Sweep Points and Sweep Type. |
Related commands
:SENSe:CORRection:USERrange:DATA:THRU
:SENSe:CORRection:USERrange:DATA:OPEN
:SENSe:CORRection:USERrange:DATA:SHORt