:SENSe:CORRection:FULLrange:DATA:LOAD
Description
Gets the actual full-range calibration data for the load calibration. If no data is available (no calibration was performed) a Query error (nr. -400) is generated.
For more information about the calibration system please refer to Calibration.
| Syntax | Response |
|---|---|
:SENSe<ch>:CORRection:FULLrange:DATA:LOAD? |
<numeric array 1>,<numeric array 2> |
| Parameters | accepted format | description |
|---|---|---|
<ch> |
{ 1 | empty } |
Currently only one channel is supported. Empty means 1 is used as default. |
| Response value | Description |
|---|---|
<numeric array 1> |
List of the real part of the calibration values separated by a comma (,) (the amount of elements is equal to the number of points defined in :SENSe:CORRection:FULLrange:DATA:FREQuencies) |
<numeric array 2> |
List of the imaginary part of the calibration values separated by a comma (,) (the amount of elements is equal to the number of points defined in :SENSe:CORRection:FULLrange:DATA:FREQuencies) |
Related commands
:SENSe:CORRection:FULLrange:DATA:THRU
:SENSe:CORRection:FULLrange:DATA:OPEN
:SENSe:CORRection:FULLrange:DATA:SHORt