:STATus:QUEStionable:PTRansition
Description
Gets or sets the value of the positive transition filter register. Setting a bit in the positive transition filter register causes a 0
to 1
transition of the corresponding questionable condition register bit to set the associated bit in the questionable event register.
Note
Reading the register has no effect on the content of the register.
Syntax | Parameter |
---|---|
:STATus:QUEStionable:PTRansition |
<numeric > |
Parameters | accepted format | possible Errors |
---|---|---|
{ <numeric > | MINimum | MAXimum } |
Range from 0 to 65535 |
Numeric data error |
Note
Negative parameter values are treated as two's complement. Positive parameters exceeding the predefined range are bitwise AND masked with the maximal allowed value.
Bit value | Code | Description |
---|---|---|
Bit 0 - 1 | VOLTage |
The quality of the voltage measurement can not be guaranteed. This bit is set when an overload was detected during measurement. |
Bit 1 - 2 | CURRent |
This bit is not used. |
Bit 2 - 4 | TIME |
This bit is not used. |
Bit 3 - 8 | POWer |
This bit is not used. |
Bit 4 - 16 | TEMPerature |
This bit is not used. |
Bit 5 - 32 | FREQuency |
This bit is not used. |
Bit 6 - 64 | PHASe |
This bit is not used. |
Bit 7 - 128 | MODulation |
This bit is not used. |
Bit 8 - 256 | CALIbration |
The quality of the calibration can not be guaranteed. This bit is set when the frequency range of an active full range calibration is not covering the frequency range of a performed measurement. |
Bit 9 - 512 | Not used |
This bit is not used. |
Bit 10 - 1024 | Not used |
This bit is not used. |
Bit 11 - 2048 | Not used |
This bit is not used. |
Bit 12 - 4096 | Not used |
This bit is not used. |
Bit 13 - 8192 | INSTrument Summary |
A summary of the instruments register. |
Bit 14 - 16384 | Command Warning |
This bit is not used. |
Bit 15 - 32768 | Not used |
The usage of the most significant bit is not allowed since some controllers may have difficulty reading a 16 bit unsigned integer. |
Sample usage
Enable all positive transitions to be reported in the associated questionable event register.
Input:
var visaSession = new TcpipSocket(VISA_RESOURCENAME); visaSession.RawIO.Write(":STAT:QUES:PTR 65535\n");
Related commands
:STATus:QUEStionable:CONDition
:STATus:QUEStionable:NTRansition