:SENSe:FREQuency:STOP
Description
This command gets or sets the stop frequency of a sweep. If the stop frequency is altered, the center frequency the frequency span and the start frequency are changed accordingly.
If the specified parameter is out of the allowable setup range, the minimum value (if the lower limit of the range is not reached) or the maximum value (if the upper limit of the range is exceeded) is set.
Syntax | Parameter |
---|---|
:SENSe<ch>:FREQuency:STOP |
{ <numeric > | MINimum | MAXimum } |
Parameters | accepted format | allowed suffixes | possible Errors | description |
---|---|---|---|---|
<ch > |
{ 1 | empty } |
none |
Invalid channel index |
Currently only one channel is supported. Empty means 1 is used as default. |
{ <numeric > | MINimum | MAXimum } |
Range from Min. frequency to Max. frequency | Hz |
Numeric data error |
The frequency value, at which the sweep should stop. |
Note
MHZ
is interpreted as Milli Hz
(1E-3) and MAHZ
is interpreted as Mega Hertz
(1E6).
Presets
For presets please refer to Measurement Modes.
Sample Usage
Configure and start a sweep measurement with start frequency of 1kHz
and stop frequency of 4MHz
.
Input:
var visaSession = new TcpipSocket(VISA_RESOURCENAME); visaSession.RawIO.Write(":SENS:FREQ:STAR 1kHz\n"); visaSession.RawIO.Write(":SENS:FREQ:STOP 4MAHz\n"); visaSession.RawIO.Write(":TRIG:SING\n"); // wait for the measurement to finish visaSession.RawIO.Write("*OPC?\n"); visaSession.RawIO.ReadString(); visaSession.RawIO.Write(":SENS:FREQ:DATA?\n"); var frequencyValues = visaSession.RawIO.ReadString(); visaSession.RawIO.Write(":CALC:DATA:SDAT?\n"); var measurementValues = visaSession.RawIO.ReadString();
Related commands
Related commands in the Automation Interface
ConfigureSweep(Double, Double, Int32, SweepMode) (Gain)
ConfigureSweep(Double, Double, Int32, SweepMode) (OnePort)
ConfigureSweep(Double, Double, Int32, SweepMode) (S11)
ConfigureSweep(Double, Double, Int32, SweepMode) (S21)
ConfigureSweep(Double, Double, Int32, SweepMode) (SeriesThru)
ConfigureSweep(Double, Double, Int32, SweepMode) (ShuntThru)
StopFrequency (Gain)
StopFrequency (OnePort)
StopFrequency (S11)
StopFrequency (S21)
StopFrequency (SeriesThru)
StopFrequency (ShuntThru)