:STATus:OPERation:NTRansition
Description
Gets or sets the value of the negative transition filter register. Setting a bit in the negative transition filter register causes a 1
to 0
transition of the corresponding operation condition register bit to set the associated bit in the operation event register.
Note
Reading the register has no effect on the content of the register.
Syntax | Parameter |
---|---|
:STATus:OPERation:NTRansition |
<numeric > |
Parameters | accepted format | possible Errors |
---|---|---|
{ <numeric > | MINimum | MAXimum } |
Range from 0 to 65535 |
Numeric data error |
Note
Negative parameter values are treated as two's complement. Positive parameters exceeding the predefined range are bitwise AND masked with the maximal allowed value.
Bit value | Code | Description |
---|---|---|
Bit 0 - 1 | CALIbrating |
The device is calibrating. |
Bit 1 - 2 | SETTling |
This bit is not used. |
Bit 2 - 4 | RANGing |
This bit is not used. |
Bit 3 - 8 | SWEeping |
The device is performing a sweep. |
Bit 4 - 16 | MEASuring |
The device is performing a measurement. |
Bit 5 - 32 | Waiting for TRIGger Summary |
The device is waiting for a trigger signal. |
Bit 6 - 64 | Waiting for ARM Summary |
This bit is not used. |
Bit 7 - 128 | CORRecting |
This bit is not used. |
Bit 8 - 256 | Not used |
This bit is not used. |
Bit 9 - 512 | REQuesting device lock |
The device lock is requested (the internal calibration is currently running). |
Bit 10 - 1024 | LOCKing device |
The device is locked (in use). |
Bit 11 - 2048 | Not used |
This bit is not used. |
Bit 12 - 4096 | Not used |
This bit is not used. |
Bit 13 - 8192 | INSTrument Summary |
A summary of the instruments register. |
Bit 14 - 16384 | PROGram Running |
This bit is not used. |
Bit 15 - 32768 | Not used |
This bit is not used. |
Sample usage
Enable all negative transitions to be reported in the associated operation event register.
Input:
var visaSession = new TcpipSocket(VISA_RESOURCENAME); visaSession.RawIO.Write(":STAT:OPER:NTR 65535\n");