:SOURce:POWer:LIST:FREQuency
Description
Gets or sets the frequency points in Hz used for level shaping.
Syntax
:SOURce<ch>:POWer:LIST:FREQuency
| Command | Query | Query Response |
|---|---|---|
| Yes | Yes | The frequencies used for shaping in Hz: <numeric value 1>[,<numeric value 2>,...,<numeric value N> ] |
| Parameters | Accepted Format |
Allowed Suffixes |
Possible Errors |
Description |
|---|---|---|---|---|
[{ <numeric value 1> | MINimum | MAXimum },...,{ <numeric value N> | MINimum | MAXimum } ] |
List of frequencies in the range from Min. frequency to Max. frequency | Hz |
none |
The frequencies in Hz |
To learn more about channels and traces, please refer to the SCPI command structure
Note
If no values are defined the query runs into a timeout.
Note
One value is mandatory. The amplitude values and frequency values need to fit. If the number of frequency points and the number of amplitudes do not match the exceeding values will be ignored.