:MMEMory:LOAD:CORRection:FULLrange
Description
Allows to load the previously saved full-range calibration from the device when connected to a Bode 500 device or from the local file system when measuring with Bode 100.
Important
In case of Bode 100 devices the parameter must represent a valid file format and must have '.mcalx' file extension.
If no path information is provided the calibration file will be loaded from the default directory C:\ProgramData\OMICRON_Lab\BodeAnalyzer\Scpi\Calibration.
For Bode 500 only a valid file name is required (NO file path information allowed, extension not required).
Syntax
:MMEMory:LOAD:CORRection:FULLrange
| Command | Query |
|---|---|
| Yes | No |
| Parameters | Accepted Format |
Allowed Suffixes |
Possible Errors |
Description |
|---|---|---|---|---|
<string> |
string value | none |
Illegal parameter value, File name not found--256 |
The calibration file name. |
Sample usage
Initialize a S21 measurement using a Bode 100 device and load a previously saved file from the file system.
visaSession.write(':CALC:PAR:DEF S21\n')
# wait for the action to finish
visaSession.query('*OPC?\n')
visaSession.write(':MMEM:LOAD:CORR:FULL "s21calibration.mcalx"\n')
# wait for the action to finish
visaSession.query('*OPC?\n')
Related commands
:MMEMory:CATalog:CORRection
:MMEMory:DELete:CORRection
:MMEMory:LOAD:CORRection:FULLrange
:MMEMory:LOAD:CORRection:USERrange
:MMEMory:STORe:CORRection