:STATus:QUEStionable:CONDition
Description
This command returns the content of the questionable condition register.
Note
Reading the register has no effect on the content of the register.
Syntax
:STATus:QUEStionable:CONDition?
| Command | Query | Query Response |
|---|---|---|
| No | Yes | Value of the questionable condition register in range from 0 to 32767. |
| Bit value | Code | Description |
|---|---|---|
| Bit 0 - 1 | VOLTage |
The quality of the voltage measurement can not be guaranteed. This bit is set when an overload was detected during measurement. |
| Bit 1 - 2 | CURRent |
This bit is not used. |
| Bit 2 - 4 | TIME |
This bit is not used. |
| Bit 3 - 8 | POWer |
This bit is not used. |
| Bit 4 - 16 | TEMPerature |
This bit is not used. |
| Bit 5 - 32 | FREQuency |
This bit is not used. |
| Bit 6 - 64 | PHASe |
This bit is not used. |
| Bit 7 - 128 | MODulation |
This bit is not used. |
| Bit 8 - 256 | CALIbration |
The quality of the calibration can not be guaranteed. This bit is set when the frequency range of an active full range calibration is not covering the frequency range of a performed measurement. For more information about the calibration system please refer to Calibration. |
| Bit 9 - 512 | Not used |
This bit is not used. |
| Bit 10 - 1024 | Not used |
This bit is not used. |
| Bit 11 - 2048 | Not used |
This bit is not used. |
| Bit 12 - 4096 | Not used |
This bit is not used. |
| Bit 13 - 8192 | INSTrument Summary |
A summary of the instruments register. |
| Bit 14 - 16384 | Command Warning |
This bit is not used. |
| Bit 15 - 32768 | Not used |
The usage of the most significant bit is not allowed since some controllers may have difficulty reading a 16 bit unsigned integer. |
Sample usage
response = visaSession.query(':STAT:QUES:COND?\n')
print(response)
Related commands
:STATus:QUEStionable[:EVENt]?
:STATus:QUEStionable:ENABle
:STATus:QUEStionable:NTRansition
:STATus:QUEStionable:PTRansition