    :OUTPut:AOFF

    Description

    Selects, whether the power source should be automatically turned off after a single sweep cycle. This is useful, if only a single sweep is performed. The DUT does not heat up unnecessarily. When performing a continuous measurement, it is considered to deactivate the automatic output power off feature, because otherwise the source is powered down between each sweep run.

    Note

    The source state will NOT be changed immediately when value is applied!

    Syntax

    :OUTPut:AOFF

    Command Query
    Query Response
    Yes Yes Boolean value indicating whether the Automatic Output Power OFF Feature is activated or deactivated.
    Parameters
    Accepted Format
    Allowed Suffixes
    Possible Errors
    Description
    <bool> { ON | OFF | 1 | 0 } none Illegal parameter value Boolean value indicating whether the Automatic Output Power OFF Feature should be activated or deactivated.

    Options

    Options Description
    OFF | 0 Automatic Output Power OFF Feature deactivated.
    ON | 1 Automatic Output Power OFF Feature activated.

    Sample usage

    Enable and start a continuous measurement and ensure that the power source remains active between each sweep run.

    visaSession.write(':TRIG:SOUR INT\n')
    
    # ensure that the power source stays activate between each sweep run
    visaSession.write(':OUTP:AOFF OFF\n')
    visaSession.write(':INIT:CONT ON\n')
    