    :INPut:IMPedance:CH2

    Description

    This command gets or sets the input impedance of channel 2.

    Syntax

    :INPut<ch>:IMPedance<tr>:CH2

    Command Query
    Query Response
    Yes Yes The termination impedance (either 50Ω or 1MΩ)
    Parameters
    Accepted Format
    Allowed Suffixes Possible Errors Description
    { 50[Ohm] | 1e6[Ohm] } numeric value (either 50 or 1MA) Ohm Settings conflict The termination impedance.

    To learn more about channels and traces, please refer to the SCPI command structure

    Warning

    Only in certain measurement modes certain impedance values can be changed. See Allowed Changes for further details.

    Allowed Changes

    measurement type CH1 CH2
    GainPhase yes yes
    S11 no yes
    S21 no yes
    Z - ECOupler yes yes
    Z - EBRidge yes yes
    Z - IADapter no no
    Z - P1Reflection no no
    Z - TSERies no yes
    Z - TSHunt no yes
    Z - VCGain yes yes

    Presets

    For presets please refer to Measurement Modes and Default Settings.

    Sample usage

    Set the impedance of channel 2 to 1MOhm in GainPhase measurement.

    visaSession.write(':CALC:PAR:DEF GAIN\n')
    visaSession.write(':INP:IMP:CH2 1e6\n')
    

    Related commands

    :INPut:IMPedance:CH1