:TRIGger[:SEQuence]:SOURce
Description
Selects the expected trigger source.
For more information about the trigger system, please refer to Trigger System.
| Syntax | Parameter | 
| :TRIGger[:SEQuence]:SOURce | { INTernal|BUS|EXTernal} | 
| Parameters | accepted format | allowed suffixes | possible Errors | description | 
| { INTernal|BUS|EXTernal} | exactly the given options | none | Illegal parameter value | Source of the trigger. | 
 
| Syntax | Response | 
| :TRIGger[:SEQuence]:SOURce? | { INT|BUS|EXT} | 
| Response value | description | 
| { INT|BUS|EXT} | Source of the trigger. | 
 
Options
| Options | Description | 
| BUS | The BUS trigger is selected e.g. the device can be triggered with the *TRGcommand, but also with:TRIG:SING | 
| INTernal | The Internal trigger is selected e.g. the device triggers itself. This can be useful when performing continuous measurements. | 
| EXTernal | The External trigger is selected. The availability of an external trigger depends on the device used. | 
Presets
| default value |  | 
| { INTernal|BUS|EXTernal} | INTernal | 
Sample usage
Initiate the trigger and perform a single trigger cycle.
var visaSession = new TcpipSocket(VISA_RESOURCENAME);
visaSession.RawIO.Write(":TRIG:SOUR BUS\n");
visaSession.RawIO.Write(":INIT\n");
visaSession.RawIO.Write("*TRG\n");
 
visaSession.write(':TRIG:SOUR BUS\n')
visaSession.write(':INIT\n')
visaSession.write('*TRG\n')
 
:INITiate:CONTinuous
:INITiate[:IMMediate]
References
Trigger System