:SENSe:CORRection:LOAD
Description
Gets or sets the resistance value in Ω used for the load calibration.
Note
The load value can only be set, if the selected measurement mode supports a load calibration. Otherwise a Settings Conflict
error is raised.
Syntax |
Parameter |
:SENSe<ch>:CORRection:LOAD |
<numeric > |
Parameters |
accepted format |
allowed suffixes |
possible Errors |
description |
<ch > |
{ 1 | empty } |
none |
Invalid channel index |
Currently only one channel is supported. Empty means 1 is used as default. |
{ <numeric > | MINimum | MAXimum } |
Range from 1Ω to 1kΩ |
Ohm |
Settings conflict |
The load value in Ohm. |
Syntax |
Response |
:SENSe<ch>:CORRection:LOAD? |
<numeric > |
Parameters |
accepted format |
description |
<ch > |
{ 1 | empty } |
Currently only one channel is supported. Empty means 1 is used as default. |
Response value |
description |
<numeric > |
The load value in Ohm. |
Sample usage
Configure an S11
measurement, set the load value to 100 Ohm and perform the full range load calibration.
var visaSession = new TcpipSocket(VISA_RESOURCENAME);
visaSession.RawIO.Write(":CALC:PAR:DEF S11\n");
visaSession.RawIO.Write(":SENS:CORR:LOAD 100Ohm\n");
visaSession.RawIO.Write(":SENS:CORR:FULL:LOAD\n");
visaSession.RawIO.Write("*OPC?*\n");
visaSession.RawIO.ReadString();
visaSession.write(':CALC:PAR:DEF S11\n')
visaSession.write(':SENS:CORR:SDEL 100Ohm\n')
visaSession.write(':SENS:CORR:FULL:LOAD\n')
visaSession.query('*OPC?\n')